Jamac Metering

Calibration & Testing Laboratory

Our Scope

Our Laboratory Scope of Accreditation​

The laboratory has gone through rigorous inspection by accreditation body, and it has demonstrated its technical competence to operate in accordance with MS ISO/IEC 17025:2017.

Scope of testing: Electrical (Laboratory & Site)

Materials / Products TestedType of Test / Properties Measured / Range of MeasurementStandard Test Method / Equipment / Techniques
Single Phase & Three Phase Energy Meter
(Electromechanical)



Active Energy Class 1 & 2
Limit of error due to variation of currentMS/IEC 62053-11:2009 clause 8.1
No load conditions test*MS/IEC 62053-11:2009 clause 8.3.1
Starting test*MS/IEC 62053-11:2009 clause 8.3.2
Meter constantMS/IEC 62053-11:2009 clause 8.4
Single Phase & Three Phase Energy Meter
(Static)



Active Energy Class 0.2s, 0.5s, 1 & 2
Reactive Energy Class 2 & 3
Limit of error due to variation of currentMS/IEC 62053-21:2009 clause 8.1

MS/IEC 62053-22:2009 clause 8.1

MS/IEC 62053-23:2009 clause 8.1
No load conditions test*MS/IEC 62053-21:2009 clause 8.3.2

MS/IEC 62053-22:2009 clause 8.3.2

MS/IEC 62053-23:2009 clause 8.3.2
Starting test*MS/IEC 62053-21:2009 clause 8.3.3

MS/IEC 62053-22:2009 clause 8.3.3

MS/IEC 62053-23:2009 clause 8.3.3
Meter constantMS/IEC 62053-21:2009 clause 8.3.4

MS/IEC 62053-22:2009 clause 8.3.4

MS/IEC 62053-23:2009 clause 8.3.4
* Only applicable for Laboratory testing.

Scope of calibration: Electrical

Instrument Calibrated / Measurement ParameterRangeCalibration and Measurement Capability Expressed as an Uncertainty (±)*Remarks
Single Phase & Three Phase Energy Meter
(Electromechanical & Static)



Active Energy Classes 1 & 2
Energy Measurement Error at input levels:
Voltage:57 Vto 276 V
(Vp-n)

Current: 0.01 A to 100 A

Power Factor:
0, 0.50 (Lag and Lead),
0.866 (Lag and Lead),
1.00 (Unity)

Frequency: 50 Hz






0.04%
In-house calibration procedure with reference to standards:


MS/IEC 62053-11:2009

MS/IEC 62053-21:2009

MS/IEC 62053-22:2009

MS/IEC 62053-23:2009
Single Phase & Three Phase Energy Meter
(Electromechanical & Static)

Active Energy Class 0.2s, 0.5s, 1 & 2
Reactive Energy Class 2 & 3




0.04%
0.04%

Have any enquiry regarding to the energy meter calibration and testing?

Scroll to Top